The Ikus driving unit is designed for stationary use with the EXCOR® Adult and EXCOR® Pediatric blood pumps.

  • Pneumatic pump supply system for pulsatile blood flow
  • Flexible: for all pump sizes, driving pressures at rates from 30 to 150 bpm
  • Triple redundancy of core components for maximum safety
  • Up to 30 minutes battery operation for patient mobility within the hospital
Stationary driving unit Ikus
Article no.
D03I – 001 EXCOR® Ikus stationary driving
unit, 230 V, 50 HZ
D03I – 002 EXCOR® Ikus stationary driving
unit, 220 V, 60 HZ
D03I – 110 EXCOR® Ikus stationary driving
unit, 115 V, 60 HZ
D03I – 112 EXCOR® Ikus stationary driving
unit, 100 V, 50 Hz
D03I – 113 EXCOR® Ikus stationary driving
unit, 100 V, 60 Hz
1600190
1600422
Tank unit
Features Description
For use with pump
sizes (univentricular/
biventricular)
10, 15, 25, 30, 50, 60, 80 ml***
Systolic pressure 60 to 350 mmHg
Diastolic pressure -100 to 0 mmHg
Pump frequency 30 to 150 bpm
Relative systolic duration 20 to 70%
Battery operation min. 30 minutes.
Alarm after 30 minutes (at max.
30 °C ambient temperature,
otherwise less)
Battery charge duration 6 h
Maintenance interval 2000 operating hours or 6
months
(at max. 30 °C ambient
temperature, otherwise 1000
operating hours or 3 months)
Ambient temperature,
operation
+10 to +30 °C with restrictions
up to +35 °C
Ambient temperature,
transportation and storage
-10 to +50 °C
6 h rest phase after transport
before commissioning
Relative ambient air
humidity
45 to 75 %

***cannot be supported with D03I-112 & D031-113